Skip to main content
Thermo Fisher Scientific

Troubleshoot intermittent AS-AP errors occurring immediately after an injection

Use this article to troubleshoot a Dionex AS-AP Autosampler that intermittently reports an error immediately after an injection and may operate normally again after restarting. Use the Chromeleon Audit Trail, AS-AP self-test, and individual component commands to determine whether the fault originates from the injection valve, syringe drive, X/Z needle drive, carousel, motor-controller electronics, valve I2C communication, or another common autosampler component before replacing parts.

Issue

The Dionex AS-AP Autosampler intermittently reports an error immediately after an injection.

The autosampler may operate normally again after restarting or power cycling. Because several AS-AP operations can occur near the end of an injection cycle, the component that reports the error may not be the injection valve itself.

Possible affected functions include:

  • Injection valve
  • Syringe drive
  • Sampling needle / Z drive
  • Horizontal / X drive
  • Carousel
  • Motor-controller electronics
  • Optional-valve communication
  • Common autosampler electronics or communication paths

Symptom pattern:

  • Error occurs intermittently.
  • Error commonly appears immediately after an injection or during the transition to the next autosampler operation.
  • Restarting or power cycling may temporarily restore operation.
  • The exact failed component may not be obvious from the timing of the error alone.

Environment

  • Product: Thermo Scientific Dionex AS-AP Autosampler
  • Software: Thermo Scientific Dionex Chromeleon CDS

Resolution

Recommended Troubleshooting Steps

  • Record the exact Chromeleon Audit Trail error text before restarting the autosampler.

    Do not troubleshoot from the general description "autosampler error" or from the timing of the error alone.

    Record:

    • Complete error wording
    • Error number, if present
    • Time of the error
    • Command or event immediately preceding the error
    • Whether the AS-AP remains connected and responsive in Chromeleon
  • Determine what the AS-AP was physically doing when the error occurred.

    Identify whether the error occurred while the autosampler was:

    • Switching an injection or optional valve
    • Moving the syringe
    • Retracting or lowering the needle
    • Moving the needle horizontally
    • Moving the carousel
    • Beginning an injection wash
    • Returning to a home or endpoint position

    An error displayed immediately after an injection does not prove the injection valve is defective. The failing operation may be the next motion command.

  • Run the AS-AP Reset / Self Test.

    Record whether the test:

    • Passes completely
    • Fails consistently on one subsystem
    • Fails intermittently
    • Produces a new or more specific Audit Trail error

    Current AS-AP service documentation uses the Audit Trail and self-test as primary tools for isolating AS-AP faults. 065437_AS-AP_Service Manual_Rev…

  • If the error identifies the injection or optional valve, exercise the valve independently.

    From Chromeleon, manually switch the valve repeatedly between its available positions.

    Observe whether:

    • The valve reaches both positions consistently.
    • The error occurs at one particular valve position.
    • The valve fails outside of an analytical injection.
    • The valve operates reliably during manual testing but fails only during a sequence.

    If the valve fails independently of injections, investigate the valve mechanics, valve electronics, cabling, and endpoint operation before investigating unrelated AS-AP drives.

  • Test the syringe drive independently.

    Run repeated syringe operations such as:

    • Prime
    • Wash
    • Draw
    • Dispense

    Observe the complete syringe stroke.

    If the error repeatedly occurs at the same syringe position or direction change, investigate the syringe drive, syringe mechanics, syringe motor controller, and related connections.

  • Test the Z/needle drive independently.

    Repeatedly command vertical needle movement.

    Observe whether the needle:

    • Reaches the commanded position
    • Returns completely
    • Hesitates or stalls
    • Produces an endpoint or alignment error

    If only the Z movement fails, keep the investigation within the Z-drive/needle/mechanical/controller branch.

  • Test the X drive independently.

    Repeatedly move the needle horizontally between appropriate autosampler positions, such as:

    • Sample position
    • Wash port
    • Injection port

    If horizontal movement fails while Z movement remains reliable, investigate the X-drive branch rather than the entire autosampler.

  • Test the carousel independently.

    Repeatedly command movement to several different tray or vial positions.

    If carousel motion alone produces the error, investigate:

    • Carousel mechanics
    • Carousel position sensing
    • Carousel motor controller
    • Associated cabling or connections
  • Interpret endpoint-related errors by identifying the failed axis.

    For errors such as an endpoint not being reached, determine which destination or movement was commanded when the error occurred.

    Map the failure as:

    Command → physical axis/component → motor controller → cable/connector → position or home sensing → mechanics

    Do not replace the main board solely because an endpoint error is intermittent.

  • Differentiate optional-valve I2C faults from motor-drive faults.

    The AS-AP service architecture identifies I2C communication for the optional-valve connection. The service manual identifies the valve connection with 5 V, I2C Data, I2C Clock, and Ground. 065437_AS-AP_Service Manual_Rev…

    If the error follows an optional-valve command, investigate the valve electronics and associated I2C communication path.

    If the error follows syringe, X, Z, or carousel movement, investigate the applicable motor-control branch instead.

  • Do not diagnose a generic "parallel bus" failure without supporting error or schematic evidence.

    Current AS-AP service documentation explicitly identifies motor-controller signals and optional-valve I2C communication. Do not assign an undocumented bus failure merely because the symptom is intermittent.

  • Use repeated component testing to determine whether the failure is local or common.

    Interpret the results as follows:

    • Only the valve fails: valve/mechanical/valve-electronics branch increases in probability.
    • Only one motion axis fails: that drive, controller, sensor, cable, or mechanical assembly increases in probability.
    • Several independent drives fail intermittently: common power, communication, main-board, or shared connection faults increase in probability.
    • All manual component tests pass but sequence operation fails: investigate sequence command order, timing, state transitions, and software/communication behavior.
    • Restart restores several otherwise unrelated functions: common electronics or controller-state problems become more likely, but the restart alone does not identify the failed part.
  • If the fault remains intermittent, run a controlled reproduction sequence.

    Run repeated identical injections while recording:

    • Audit Trail
    • Injection number
    • Valve state
    • Syringe activity
    • Needle/X/Z activity
    • Carousel activity
    • Wash commands
    • Exact failure point

    A repeated sequence is preferable to replacing components based on a single intermittent event.

  • Inspect or reseat connectors only after the failing subsystem has been narrowed.

    If testing identifies a specific motor controller, valve path, or common electronic branch, inspect the corresponding:

    • Cable
    • Connector
    • Ground
    • Motor-controller connection
    • Valve electronics connection

    Avoid disconnecting or reseating every connection in the AS-AP without first determining which branch is implicated.

  • Replace a component only when the isolation test supports it.

    After repair, repeat:

    • Self-test
    • Manual component exercise
    • Multiple analytical injections under the original customer method

    Verify that the original intermittent error no longer occurs.

Analysis