Troubleshoot no chromatographic peaks on a Dionex Aquion IC system
Issue
The Dionex Aquion or Aquion RFIC system appears to be running, but no chromatographic peaks are observed after sample or standard injections.
The system may have apparently normal pump operation, pressure, and baseline conductivity, but injections produce no detectable analyte peaks.
Environment
- Product: Thermo Scientific Dionex Aquion IC
- Product: Thermo Scientific Dionex Aquion RFIC
- Software: Thermo Scientific Chromeleon 7 CDS
Resolution
Before troubleshooting, record:
- Exact Aquion configuration
- Autosampler or manual injection configuration
- Isocratic or gradient method
- Eluent type and concentration
- Method flow rate
- Actual system pressure
- Background conductivity
- Column and guard
- Suppressor type and operating mode
- Injection loop volume
- Chromeleon version
- Exact Audit Trail messages
- Whether the problem affects all injections
- Whether the failure was sudden, progressive, intermittent, or began after maintenance or a method/configuration change
Recommended Troubleshooting Steps
-
Use a fresh known standard.
Use a stable standard at a concentration that normally produces an obvious response. Do not begin troubleshooting with an unknown customer sample.
Confirm that the vial contains the correct standard and that the expected injection volume and method are being used.
-
Determine whether the sample-introduction system is loading the injection loop.
If an autosampler is installed, perform an appropriate manual fixed-loop injection when practical.
Manually fill the loop using the correct injection arrangement and sufficient sample volume to ensure the loop is completely filled.
Run the method and observe the raw conductivity signal.
- If manual injection produces normal peaks, move troubleshooting upstream toward the autosampler, sample-transfer path, vial, needle, delivery settings, or autosampler plumbing.
- If manual injection also produces no peaks, continue troubleshooting the Aquion injection valve and downstream chromatograph.
-
Verify pump pressure and actual eluent flow.
Record the system pressure at the method flow rate and compare it with the normal value for the installed column set and method.
Measure actual flow independently by collecting the detector or appropriate outlet flow for a defined period or by using an appropriate calibrated flow measurement device.
Compare measured flow with commanded flow.
- Very low pressure or low collected flow can indicate inlet starvation, air, loss of prime, leakage, or another hydraulic delivery problem.
- Unexpectedly high pressure suggests a restriction in the high-pressure flow path.
- Normal, stable pressure and correct measured flow make a major pump-delivery failure less likely.
Do not change pump calibration to compensate for an unresolved hydraulic problem.
-
Verify injection-valve operation and loop plumbing.
Use Chromeleon direct control or the applicable instrument controls to command the injection valve between Load and Inject.
Confirm that:
- The valve changes position when commanded.
- The installed sample loop is correctly connected.
- Tubing is connected to the proper valve ports.
- The sample loop is not blocked.
- There is no visible leakage.
- The sample is not being routed directly to waste instead of into the analytical flow path.
Repeat a known-standard injection after confirming valve operation.
-
Simplify the flow path to separate injection/detection from separation and suppression.
When the applicable Aquion procedure and available test hardware permit, temporarily simplify the conductivity flow path so that the injection valve can be evaluated with the separation and suppression components removed from the test.
Thermo Fisher's integrated-IC OQ approach uses this type of configuration by bypassing applicable column, suppressor, and related devices and connecting an appropriate test path from the injection valve toward the conductivity detector. This is a useful diagnostic principle for determining whether an injected standard can reach and be detected by the conductivity cell. [Procedures]CDS OQ Chromeleon A…
- If a peak appears in the simplified path, the fault is likely within one or more of the components that were bypassed.
- If no peak appears, continue concentrating on the injection valve, sample loop, conductivity detector, and Chromeleon signal/configuration.
Use only plumbing and pressure conditions supported for the Aquion configuration. Do not substitute Inuvion or Integrion service plumbing by analogy.
-
Evaluate the column and guard.
Reintroduce the guard and analytical column only after the simplified path is understood.
Verify:
- Correct column and guard chemistry
- Correct flow direction
- Correct tubing connections
- Appropriate column conditioning/equilibration
- Normal pressure contribution
- No obvious restriction
- No recent column storage, contamination, or method changes that could explain the failure
A severe restriction generally produces increased pressure. If pressure remains normal, a completely blocked column is less likely, although contamination, incorrect chemistry, or improper plumbing can still prevent usable chromatography.
-
Evaluate the suppressor separately.
For suppressed-conductivity methods, verify:
- Correct suppressor type for the application
- Correct operating mode
- Suppressor is enabled where required
- Appropriate regenerant or external-water operation where applicable
- No bubbles or obvious leakage
- Correct suppressor plumbing
- Background conductivity is reasonable for the application
If the baseline is very high, unstable, or pegged, suppression or detector problems move higher in the fault tree.
If the baseline is normal and stable but no injected response appears, first prove that sample is reaching the detector before replacing the suppressor.
-
Check conductivity-detector response independently of chromatography.
Establish a clean detector baseline with the appropriate fluid flowing through the conductivity cell.
Thermo Fisher integrated-IC qualification procedures evaluate conductivity detector baseline performance by flowing water through the detector cell and allowing the signal to equilibrate before noise and drift are evaluated. [Procedures]CDS OQ Chromeleon A…
Observe:
- Conductivity value
- Baseline stability
- Noise
- Drift
- Bubbles in the flow path
- Detector temperature behavior where applicable
- Whether the signal changes normally rather than remaining frozen
An abnormal detector baseline shifts troubleshooting toward water quality, bubbles, cell condition, temperature control, detector configuration, or detector hardware.
-
Verify Chromeleon instrument configuration and acquired signal.
Confirm that Chromeleon is configured for the actual Aquion detector and that the correct conductivity channel is being acquired.
Check:
- Instrument Configuration
- Detector device and signal
- Instrument Method
- Acquisition channel
- Injection-valve commands
- Sequence instrument assignment
- Audit Trail
- Device communication
- Raw conductivity trace
Thermo Fisher's integrated-IC OQ material identifies Aquion conductivity configurations using the applicable conductivity signal and emphasizes correct instrument-method and signal configuration before evaluating detector performance. [Procedures]CDS OQ Chromeleon A…
-
Inspect the raw chromatogram before diagnosing an instrument failure.
Determine whether there is truly no detector response or whether a response exists but is not being identified or integrated.
- If the raw signal contains a peak but Chromeleon does not report it, troubleshoot processing, retention windows, peak detection, integration, or quantification settings.
- If the raw signal shows no response at the injection time, continue physical sample-path and detector troubleshooting.
-
Return the system to the original method and verify the repair.
After identifying and correcting the fault, reinstall the normal column, guard, suppressor, autosampler, and plumbing as applicable.
Verify with the original method:
- Normal pressure
- Correct actual flow
- Stable background conductivity
- Expected retention times
- Normal peak response
- Acceptable peak shape
- Injection repeatability
- No leaks
- No relevant Audit Trail errors
Analysis