Skip to main content
Thermo Fisher Scientific

Troubleshooting AS-AP Autosampler Diverter Valve Recognition Issue with Power and Green Light Blinking

Issue

The AS-AP Autosampler with cooling is experiencing a problem where the diverter valve (DV) is not recognized. Despite the power and green light for the DV being on and consistently blinking, the autosampler is unable to configure or connect to the DVC (Diverter Valve Controller). This issue persists even after replacing the diverter valve assembly, connecting the autosampler to a laptop, and reseating DVC connections in the mainboard. The old instrument configuration file allows for DVC configuration, but errors occur when attempting to control it. Additionally, the AS-AP throws I2C bus errors when using the old configuration with the DVC .

Environment

Thermo Fisher Scientific ICS AS-AP autosampler with cooling 

Resolution

Before ordering a replacement mainboard, it is recommended to perform a self-test and power cycle the autosampler.   If the issue persists, then:

  1. Order a replacement mainboard (PN AAA-069800)  for the AS-AP Autosampler.
  2. Once the replacement mainboard is received, follow these steps:
    1.  Power off the AS-AP Autosampler and disconnect all cables.
    2. Swap/install the new AS-AP main CPU PCB (including moving the EEPROM).
    3. Reconnect all cables and power on the AS-AP Autosampler.
    4. Perform a self-test to ensure successful power-up.
    5. Check if the diverter valve is now recognized in the instrument configuration.
  3. If the issue persists after following these steps, additional troubleshooting may be necessary. Check the continuity of the DV "comm" ribbon cable traces to ensure they are functioning properly. If the AS-AP Autosampler is still in the same condition as when the Field Service Engineer arrived, return the plumbing to a single system set-up.

Helpful references:

AS-AP I2C bus.pdf

065437 - Rev 02 - ICS Series AS-AP Autosampler Service Manual

 

Analysis

It is suspected that the mainboard may be faulty and no longer able to recognize the DVC.