Troubleshooting Retention Time Issues with the NGES Suppressor
Learn how to troubleshoot retention time issues with the NGES Next Generation Electrolytic Suppressor. This guide covers common problems, steps to resolve them, and important considerations for maintaining optimal performance.
Issue
Users may experience increased retention times when using the NGES Next Generation Electrolytic Suppressor. This could be due to incorrect system setup or flow direction.
Environment
- Inuvion IC
- Aquion IC
- ICS-6000
- ICS-5000+
- NGES Suppressor
Resolution
- Verify Flow Direction:
- Ensure the suppressor is installed according to the correct direction of flow specified in the manual.
- Do not reverse the flow direction as it may cause retention time issues.
- Check System Setup:
- Confirm that the system was properly set up from the beginning. Incorrect setup can lead to retention time discrepancies.
- Review the operator manual for detailed setup instructions.
- Update Software:
- Ensure that the system is running the latest software version (minimum version 7.2.10 or 7.3.2 for compatibility).
- Update software if necessary to maintain compatibility and performance.
- Request Data:
- If issues persist, request detailed data from the customer to analyze system parameters.
- Look for changes in retention time and other performance indicators.
- Consult Support:
- Contact Level 2 support with the serial number and ticket number of the case for further assistance.
- Provide all relevant data and observations for a thorough investigation.
Cause
Increased retention time may result from incorrect flow direction or improper system setup. Ensuring compliance with setup instructions and software requirements is crucial.