Turn on AS-AP condensation pump for troubleshooting purposes
In service mode, service level use "cmd_drain_pump" to turn on the AS-AP drain/condensation pump for troubleshooting.
Issue
Lots of condensation in and under the sample trays of the AS-AP carousel. We need to determine if the condensation/drain pump is working properly.
Environment
AS-AP autosampler with temperature control
Resolution
For troubleshooting purposes, execute the command: cmd_drain_pump. Must be in service mode in Chromeleon, Service level in F8/Commands.

Click on the button by the command to start the drain pump. Click it again to turn it off. If left to run, the drain pump will run for approximately one minute.