Virtual Channels for Monitoring Electrolytics on Inuvion

Virtual Channels for Monitoring Electrolytics on Inuvion

Issue

Customer reports inconsistent chromatography results during operation. The inconsistency impacts data reliability. A Field Service Engineer requires visibility into electrolytic performance by monitoring current (mA) and voltage (V) for key components, including the suppressor, CR-TC, and EGC, to support troubleshooting and system assessment.

Environment

  • Inuvion Ion Chromatography System
  • Electrolytic components including suppressor, CR-TC, and EGC
  • Instrument control software with method scripting capability

Note: With minor script modifications, this approach may be applicable to other ion chromatography systems.

Resolution

  1. Open the instrument method in Chromeleon for the Inuvion system.
  2. Navigate to the script or method editor.
  3. Add virtual channels at the start of the run to monitor electrolytic parameters.
  4. Insert the following script:
0.000 Start Run

VirtualChannel "CR-TC Current",
Inuvion.Electrolytics.CR_TC.Current,
Type=Digital,
Unit="CR_TC.Current",
Evaluate=Yes

VirtualChannel "CR-TC Voltage",
Inuvion.Electrolytics.CR_TC.Voltage,
Type=Digital,
Unit="CT-TC.V",
Evaluate=Yes

VirtualChannel "Suppressor Current",
Inuvion.Electrolytics.Suppressor.Current,
Type=Digital,
Unit="Sup.Current",
Evaluate=Yes

VirtualChannel "Suppressor Voltage",
Inuvion.Electrolytics.Suppressor.Voltage,
Type=Digital,
Unit="Sup.V",
Evaluate=Yes

VirtualChannel "EGC Current",
Inuvion.Electrolytics.Eluent_Generator.EgcCurrent,
Type=Digital,
Unit="EGC_Current",
Evaluate=Yes

VirtualChannel "EG Voltage",
Inuvion.Electrolytics.Eluent_Generator.EgcVoltage,
Type=Digital,
Unit="EGC_Voltage",
Evaluate=Yes
  1. Save the instrument method.
  2. Run the method and confirm that virtual channels display current and voltage values.
  3. Review the monitored values to evaluate electrolytic performance and troubleshoot inconsistent chromatography.

clipboard_e2505f57759d6a2437bffdf2c89e19b71.png

Analysis

Monitoring electrolytic current and voltage provides real-time insight into component performance. If these values drop to zero or negative values during the run, chromatography will degrade.

Cause

Inconsistent chromatography may be caused by electrolytics readbacks dropping to zero (or negative values). The information from the virtual channels coupled with testing electrolytics with a test box and multimeter can help diagnose the faulty part(s).

 