What is the part number for the AS-DV sampling tip set screw?

This article provides instructions on replacing the AS-DV sampling tip assembly, including part numbers and descriptions for the sampling tip assembly and sampling tip set screw.

Issue

How to replace the sampling tip assembly for an AS-DV Autosampler?

What is the part number for the AS-DV Autosampler sampling tip set screw?

Note:  The AS-DV Autosampler sampling set screw may also be referred to as the sampling head set screw, sampling tip set screw, needle tip set screw or set screw.

Environment

AS-DV Autosampler

Part Number

The AS-DV Autosampler sampling tip set screw part number is 60-046253 and the SAP description for this is:  "SCR,M04 X0.70X014,CAP,SOC,SST"

Resolution

The following steps are used to replace the AS-DV Autosampler sampling tip assembly (Taken from the AS-DV Autosampler Service Manual):

  1. Disconnect the Sample Out Line:

    • Disconnect the sample out line from the injection valve or optional high-pressure valve and remove the fitting.
  2. Retract the Sampling Tip:

    • Press the Carousel Release button to retract the sampling tip to its fully up position.
  3. Turn Off the Power:

    • Turn off the AS-DV power and disconnect the power cord. Raise the top cover.
  4. Remove the Front Cover:

    • Remove the AS-DV front cover by gripping the indentations at each side and pulling the cover straight off toward you.
  5. Remove the Sampling Head Tower Cover:

    • Use a #1 Phillips screwdriver to remove the three flathead screws from the sampling head tower cover and lift up the cover to remove it (see Figure 5-7 in the AS-DV Autosampler Service Manual).
  6. Remove the Set Screw:

  7. Remove the Old Sampling Tip and Transfer Line:

    • Pull out and discard the old sampling tip and transfer line.
  8. Install the New Sampling Tip and Transfer Line:

    • Thread the free end of the transfer line of the new sampling tip assembly (P/N 071575) up through the sampling tip shaft and screw the tip into place on the sampling tip shaft (see Figure 8 in the AS-DV Autosampler Service Manual).
    • Slide the sampling tip shaft into place under the sampling arm and slide the transfer line back into the slot in the sampling head.
    • If the yellow sleeve does not slide easily, lubricate the transfer line with a small amount of ethanol.
  9. Align the Sampling Tip:

    • Align the two flat sides of the sampling tip so that the tip fits through the opening in the sensor block (see Figure 9 in the AS-DV Autosampler Service Manual).
    • Verify that the sampling tip shaft is pushed as far as possible into the sampling head, and then tighten the set screw firmly.
  10. Secure the Transfer Line:

    • Slide the yellow sleeve onto the transfer line and position it where the transfer line aligns with the transfer line clip on the side of the sampling arm. Secure the transfer line in the clip.
  11. Thread the Transfer Line:

    • Thread the free end of the transfer line out the aperture on the front of the Dionex AS-DV. The transfer line will be connected to the valve later.
  12. Reassemble the AS-DV:

    • Reinstall the cover of the sampling head tower.
    • Reinstall the front cover and close the carousel cover.
    • Reconnect the power cord and turn on the power.
  13. Please refer to the AS-DV Autosampler Service Manual for detailed diagrams and additional instructions.