TSX and STP Series HC -86 ULT Freezers - How to Secure Door Magnet Instruction Sheet

Description: 

Door Magnetic out of alignment with the door switch, causing door open alarm with the door closed. How to secure magnet.

Environment:

ULT HC Series:

8930086A FDE50086FA HDE60086LD RDE30086LD TDE30086FV TSX40086A
8930086D FDE50086FD HDE60086LV RDE30086LV TDE30086LA TSX40086D
8930086V FDE50086FV HLE30086A RDE40086FA TDE30086LD TSX40086G
8940086A FDE50086LA HLE30086D RDE40086FD TDE30086LV TSX40086V
8940086D FDE50086LD HLE30086V RDE40086FV TDE40086FA TSX50086A
8940086V FDE50086LV HLE40086A RDE40086LA TDE40086FD TSX50086D
8950086A FDE60086FA HLE40086D RDE40086LD TDE40086FV TSX50086G
8950086D FDE60086FD HLE40086V RDE40086LV TDE40086LA TSX50086V
8950086V FDE60086FV HLE50086A RDE50086FA TDE40086LD TSX60086A
8960086A FDE60086LA HLE50086D RDE50086FD TDE40086LV TSX60086D
8960086D FDE60086LD HLE50086V RDE50086FV TDE50086FA TSX60086G
8960086V FDE60086LV HLE60086A RDE50086LA TDE50086FD TSX60086V
CAR30086FA HDE30086FA HLE60086D RDE50086LD TDE50086FV VWR30086FA
CAR30086FD HDE30086FD HLE60086V RDE50086LV TDE50086LA VWR30086FD
CAR30086LA HDE30086FV IUE30086FA RDE60086FA TDE50086LD VWR30086FV
CAR30086LD HDE30086LA IUE30086FD RDE60086FD TDE50086LV VWR30086LA
CAR40086FA HDE30086LD IUE30086FV RDE60086FV TDE60086FA VWR30086LD
CAR40086FD HDE30086LV IUE30086LA RDE60086LA TDE60086FD VWR30086LV
CAR40086LA HDE40086FA IUE30086LD RDE60086LD TDE60086FV VWR40086FA
CAR40086LD HDE40086FD IUE30086LV RDE60086LV TDE60086LA VWR40086FD
CAR60086FD HDE40086FV IUE40086FA RLE30086A TDE60086LD VWR40086FV
CAR60086LD HDE40086LA IUE40086FD RLE30086D TDE60086LV VWR40086LA
FDE30086FA HDE40086LD IUE40086FV RLE30086V TLE30086A VWR40086LD
FDE30086FD HDE40086LV IUE40086LA RLE40086A TLE30086D VWR40086LV
FDE30086FV HDE50086FA IUE40086LD RLE40086D TLE30086V VWR60086FD
FDE30086LA HDE50086FD IUE40086LV RLE40086V TLE40086A VWR60086FV
FDE30086LD HDE50086FV IUE60086FD RLE50086A TLE40086D VWR60086LD
FDE30086LV HDE50086LA IUE60086FV RLE50086D TLE40086V VWR60086LV
FDE40086FA HDE50086LD IUE60086LD RLE50086V TLE50086A  
FDE40086FD HDE50086LV IUE60086LV RLE60086A TLE50086D  
FDE40086FV HDE60086FA RDE30086FA RLE60086D TLE50086V  
FDE40086LA HDE60086FD RDE30086FD RLE60086V TLE60086A  
FDE40086LD HDE60086FV RDE30086FV TDE30086FA TLE60086D  
FDE40086LV HDE60086LA RDE30086LA TDE30086FD TLE60086V