User Guide - Restoring Language to English in Micropure Water system controls - Rev 0 Nov 2025

This article provides a step-by-step guide on how to change the operating language of Micropure systems from German to English. It includes instructions on unlocking service menus using a code, navigating to language settings in OEM menus, and selecting English. The guide addresses issues where the system defaults to German due to glitches or electrical problems. Following these steps will help operators restore the language setting to English.

Description: 

This procedure tells step by step how to unlock the service menus using the access code, then navigate to the language setting in the OEM menus, to change the setting to English.

 

Environment:

Micropure Water Systems:
50132366 MicroPure standard
50132368 MicroPure UF
50132370 MicroPure UV/UF
50132373 MicroPure UV

50132367 MicroPure with tank standard
50132369 MicroPure with tank UF
50132372 MicroPure with tank UV/UF
50132374 MicroPure with tank UV