Performance Verification Test Collects a Background and then Returns to the System Status Overview Window

On a Nicolet iS5 or iS20, the Performance Verification test may collect the background and then return to the System Status Overview window without completing the entire test.  This happens if a window called "PV Results Window" is already open from a previous tests.  Close all previous PV Results Windows before running the test.

Issue

A user started the Performance Verification test in OMNIC from the System Status Overview window.  The tests started and collected the first background.  As soon as the collection of the background was completed, the test stopped and returned the user to the System Status Overview window.  The test never completed and no results were recorded.

Environment

OMNIC

Nicolet iS20

Nicolet iS5

Resolution

Before starting the Performance Verification test, close any windows open with the name "PV Results Window". 

Cause

  • This behavior happens if there is still a window called "PV Results Window" left open from a previously ran Performance Verification (PV) test.
  • Each time the PV test is started, OMNIC creates a PV Results Window.
  • If more than one PV Results Window are open, when the background collection is finished and OMNIC tries to add the spectrum to the window, it gets confused because there are more than one window with the same name open.

 

 

 