Analyze regions (areas, lines, and points)
With a mapping session, you can measure a collection of points at once.
Mapping is not available on systems equipped with a manual stage.
A mapping session is used to analyze regions. In OMNIC™ Paradigm™ software, a region is an area, line, or point. Regions are measured using identical measurement settings to create a chemical image of the sample. You can measure one or more regions at a time. Measuring regions requires you to first capture a mosaic.
To start a mapping session, in a Microscope tab, select Start Session in the toolbar.
-
Capture a mosaic. A mosaic is a collection of pictures of the sample that are stitched together into a single large image. The mosaic serves as the main workspace for navigating the sample and identifying areas of interest.
Before you capture a mosaic, make sure that the sample is in focus and illuminated.
-
Select the Draw Mosaic tool from toolbar.
-
Click and drag on the mosaic pane to draw a mosaic region.
-
Select Capture Mosaic.
-
Measure a background spectrum, as required by your settings.

-
-
Specify regions to analyze. You can add multiple regions to a single analysis.
To Analyze Select this tool Perform these steps Areas

-
Select the area tool.
-
Click and drag on the mosaic to draw the area.
Lines (Not available when using the polarizer during IR data collection)

-
Select the line tool.
-
Click and drag on the mosaic to draw a line.
Points

-
Select the point tool.
-
Click to add a point.
Use the pan tool to select or delete areas, lines and points.
-
-
Open the Region Queue to view and refine details for all areas, lines, and points in your measurement. You can also select a background measurement to associate with each region.
-
When you are finished adding regions, click Measure Regions.