Restart Qtegra Services to Restore Instrument Control or LabBook Updates
Restart Qtegra Windows services to resolve issues where LabBook progress stops updating or instrument control is lost.
Issue
• LabBook progress does not update during scheduled runs
• Loss of instrument control
Environment
• Qtegra ISDS Software
• Windows Operating System
• Qtegra (2.10 SR7 and later for Post Process Service)
• Applicable connected instruments
Resolution
Prerequisite:
Close Qtegra, Configurator, and all related Qtegra applications before restarting services.
- Click on the
button if available in Windows taskbar or Right-click on Windows icon and select Search - Type "Services" into the Windows search box and select Run as administrator.

- Click on
to sort the Services by Description - Scroll down to the list Qtegra services

- Right-click on the Logging Service and select Stop

- A Stop Other Services popup window will appear. Click Yes to stop the other Qtegra Services

- Right-click on Acquisition Service and select Start

- The Post Process Service was added in Qtegra 2.10 SR7 (2.10.4345.39). For Qtegra 2.10 SR7 or later: Right-click the Post Process Service and select Start.

- Once all of the Qtegra services have been restarted, Running should be displayed in the Status column for all of the Qtegra services.

Expected Result:
All Qtegra services display "Running" in the Status column. - Restart Qtegra
NOTE: If this does not work, please contact technical support for further assistance.